1Collect

1Collect

The analysis begins with on-site and/or on production line collection. The flexibility of our range of RollBag bags adapts to any type of test taking (biopsies, soils, faecal matter...).

2 Sample

2 Sample

Sampling for microbiological analysis preparation. The sample is placed in its unique bag with filter (BagFilter Pipet & Roll) for blender avoiding cross contamination. The bag stands upright with a BagOpen.

3 Dilute

3 Dilute

Weigh the sample with the DiluFlow Elite 5 kg. The DiluFlow dilutor automatically dispenses the diluent according to the chosen dilution factor with ± 2 % accuracy in accordance with the standards.

4 Homogenize

4 Homogenize

The sample is homogenized in a BagMixer SW without any contact with the blender, thus avoiding cross contamination. Bacteria are quickly extracted from the sample without being destroyed, and the filter present in the bag allows to eliminate particles in the filtrate.

5 Organize

5 Organize

Homogenized samples are closed thanks to the repositionable closing sticker of the BagFilter Pipet & Roll. They are then stored in a BagRack Slide. The pipetting area, positioned at the bottom of the bag, allows an easy sampling of the filtrate. Easily open/close the pipette access with the repositionable sticker.

6 Distribute

6 Distribute

With FlexiPump, the serial distribution of culture media, agars, diluents... is accurate, fast and sterile.

7 Plate

7 Plate

Place the filtered sample in a beaker.
The easySpiral Dilute plater automatically dilutes and plates the sample on a Petri dish, allowing 4 logs of dilution plated on the same dish. The easySpiral software collects the plating data.

8 Count

8 Count

Once the incubation is done, the automatic counting is carried out with the automatic Scan range of colony counters. Click on Click on “Count” then “Validate”. The images and results are automatically saved in your computer to ensure traceability.

Types of analysis

From sample collection to microbial analysis, we offer you a wide range of products for all analyses in food, pharmaceutical, cosmetics, environmental, veterinary industries and public institutes research.

Our products enable us to respond to innovation and research projects for new concepts in microbiology.

Example of protocol in public institutes research

Collect the sample with RollBag

Collect the sample with RollBag

Dilute the sample at 1/10th with DiluFlow

Weigh 25 g of sample and dilute 1/10 with DiluFlow

Homogenize with MiniMix lab blender

Homogenize the sample in a BagMixer

Plate with easySpiral

Plate quality indicators with easySpiral

Incubate Petri dishes with ScanStation

Place the Petri dishes for incubation in ScanStation

Obtain a compliance result with ScanStation

Enumerate colonies to obtain a compliance result with ScanStation or Scan

Collect the sample with RollBag

Collect the sample with RollBag

Dilute the sample at 1/10th with DiluFlow

Weigh 20 to 50 g faecal sample and dilute with DiluFlow

Homogenize with MiniMix lab blender

Homogenize and filter the sample in a BagMixer

Plate with easySpiral

Plate the filtrate with easySpiral using the constant mode and then place the antibiotic discs before incubation.

Analysis of multi-resistance with Scan

SIR analysis and interpretation of multi-resistance according to the CASFM veto with the Automatic Scan range

In compliance with international standards

ISO 7218

MICROBIOLOGY OF FOOD AND ANIMAL FEEDING STUFFS
General requirements and guidance for microbiological examinations

ISO 4833

FOOD MICROBIOLOGY
Horizontal method for the enumeration of microorganisms

European Pharmacopoeia

The European Pharmacopoeia is a unique reference work on the quality control of medicines in the countries that are signatories to the Convention relating to its elaboration.

EUCAST CA-SFM

Following the recommendations of the WHO Expert Committee on Biological Standardisation (technical reports no. 610, 1977), the French Society of Microbiology created a Committee on Antibiogram (CA-SFM) to determine the critical values that delimit the clinical categories (previously therapeutic categories) and to propose a guide for the determination of bacterial susceptibility to antibiotics in association with EUCAST. The critical values defined for the concentrations and diameters of the inhibition zones, as well as recommendations specific to certain species or groups of antibiotics are published in this press release.

ISO 6887

MICROBIOLOGY
General guidelines for the preparation of dilutions for microbiological examination

FDA BAM

BACTERIOLOGICAL ANALYTICAL MANUAL

Application notes

  • Dispensing repeatability of 9 mL from a 2-liter bottle with FlexiPump

    Aim: Evaluate the repeatability of a volume dispensed consecutively, after a FlexiPump calibration, from a FlexiPump calibration from a 2L diluent bottle (tube weight at the bottom and bottle filled for calibration).

    Conclusion: From the results obtained in this test, we can conclude that dispensing successive doses with the FlexiPump, from a 2-liter bottle of diluent, shows excellent repeatability. There is no drift in dispensing volume as the bottle empties.

  • Dispensing repeatability of 9 mL diluent bag with FlexiPump

    Aim: Evaluate the repeatability of a consecutively dispensed volume, after instrument calibration, from a 2L diluent bag.

    Conclusion: From the results obtained in this test, we can conclude that dispensing successive doses with the FlexiPump, from a diluent bag, shows excellent repeatability. There is no drift in dispensing volume as the bag empties.

  • 50 ML dispensing test with FlexiPump Pro

    Aim: Check the dispensing accuracy of a 50 mL volume with a FlexiPump Pro peristaltic pump, with a dispensing time of less than 4/5 seconds.

    Conclusion: The FlexiPump Pro is accurate for a 50 mL dispense, under the conditions described above. Note that under these conditions, each dispense took 2.7s.

  • Validation of easySpiral/easySpiral Pro plating
    1. Plating validation easySpiral/easyeasySpiral Pro in decreasing exponential mode
      1. Aim: To study the reliability of easySpiral/easySpiral Pro plating in decreasing exponential mode (known as the Spiral method) by carrying out a comparative study with the reference method "manual surface plating" (known as the classic technique).
      2. Conclusion: Our results show a strong correlation (R2 ≥ 0.982) between plating in Spiral mode using the easySpiral and the conventional method. Also, the average difference in log CFU/mL between the two methods is 12 times less than the maximum difference, and therefore not significant.
    2. Cleaning efficiency of easySpiral/easyeasySpiral Pro
      1. Aim: Ensure that the easySpiral/easySpiral Pro stylus is cleaned between each sample run. Check the effectiveness of stylus disinfection by inoculating with sterile diluent.
      2. Conclusion: The study shows that for all six bacterial strains, cleaning in either long or short mode enables disinfection of the stylus and avoids contamination between samples. Three disinfectants were tested: ethanol 70%, bleach 1% and a formulation based on hydrogen peroxide (3%) + peracetic acid (0.08%). Other formulations with slightly different concentrations of H2O2 and APA may also be effective.
  • Scan 1200: performance evaluation

    Aim: The aim of this study was to evaluate the performance of the Scan 1200 by comparing manual and automatic counting. For optimal comparison, Petri dishes were plated and incubated in our R&D laboratory, using standard methods to reproduce normal laboratory conditions. The same technician then counted the colonies with a Scan 1200 and manually to obtain results for evaluating the accuracy of the Scan. This document also contains a study of the analysis time per dish and an estimate of the time spent by the laboratories.

    Conclusion: Tests show in a variety of ways (regression line, correlation coefficient, average Log value difference, and ISO 7218:2007) that Scan 1200:
    — Enables faster counting (up to 80% time saving).
    — Counts as well as another user (strong relationship between the 2 methods, with an average difference of 2.35% per dish).

    The Scan 1200 is an excellent tool for laboratories needing to count large numbers of dishes accurately and without wasting time. All results can be saved in specific files (called sessions) which contain all dish photos and counts, guaranteeing analysis quality and perfect traceability.

  • ScanStation in the food industry

    Aim: The aim of this study was to evaluate the performance of the ScanStation 100 by comparing the manual and automatic methods for food and milk payment analysis. For an optimal comparison, 1238 food samples, in duplicate, were carried out on a multitude of micro-organisms according to the laboratory’s reference methods. This document also contains curves showing the evolution of bacterial load over time.

    Conclusion: Interpretation of these curves shows that the number of CFUs evolves up to 15 h of incubation. Thereafter, the number of CFU remains constant. Real-time counting during incubation enables us to quickly determine the presence of contamination, for example, and thus to define corrective actions before the end of incubation.

  • ScanStation in the food industry (study conducted on Symphony and TBX media)

    Aim: The aim of this study was to evaluate the performance of the ScanStation (ISS) by comparing manual and automatic counting on the analysis of samples seeded on Symphony and TBX media.

    Conclusion: The difference in the majority of counts does not exceed the 0.3 log CFU limit. These results show no significant difference. By reading the "Time to Result" of the various microorganisms grown on Symphony and TBX media, it is possible to anticipate count results, enabling the user to define corrective action more quickly.

  • ScanStation performance (study conducted with Salmonella typhimurium and Listeria monocytogenes)

    Aim: The aim of this study was to evaluate the performance of the ScanStation (ISS) by comparing manual and automatic decomposition of pure cultures of Salmonella typhimurium and Listeria monocytogenes.

    Conclusion: The difference in the majority of counts does not exceed the 0.3 log CFU limit. These results show no significant difference. Reading the "Time to Result" for Salmonella typhimurium and Listeria monocytogenes makes it possible to anticipate count results, enabling the user to define corrective action more quickly.

  • Counting waterborne pathogens with ScanStation, a real-time incubation and counting station. Use of filtration membranes

    Aim: The objective of this study was to evaluate the performance of the ScanStation to count in real time colonies on membrane filtration. Enumeration was performed on waterborne pathogens that have been linked to healthcare-associated infections (HAIs). Bacterial suspensions were passed through filtration membranes that were deposited on Petri dish. Colonies were manually counted and results were compared with automatic counts performed by the ScanStation.

    Conclusion: ScanStation performs well in counting colonies on filtration membranes in real time. For the seven strains tested, the automatic and manual counts are similar when bacterial suspensions are filtered on white polycarbonate membranes (without grids). For best automatic counts in this case, the recommended light configuration is white background (light from below). This study shows that bacterial colonies can be efficiently counted with ScanStation.

  • Robustness tests on ScanStation

    Aim: The aim of this study is to evaluate the performance of ScanStation (ISS) by comparing manual and automatic counting of plated samples for robustness counting assessment.

    Conclusion: ScanStation’s robustness tests showed reproducible data under both intra- and inter-machine conditions.

They trust us

Logo - ANSES

Logo - CPC

Logo - CEA

Logo - CNRS

Logo - CRCM

Logo - FDA

Logo - Inserm

Logo - Institut Pasteur

Logo - Nanjing

Logo - Peking university

Logo - University

Logo - USDA

Logo Ciri

CIRI
International research centre for infectious diseases

Using the l’easySpiral Pro saves us valuable time and allows us to be more reproducible.
AQMC

AQMC Laboratory
Consumer Safety and Health

3 dilution speeds to adapt to any type of matrix with the DiluFlow Elite dilutor
Terana

TERANA Public Analysis Laboratory
Services provided by the Conseil départemental du Puy-de-Dôme, France

We have a DiluFlow Elite double pump with booster kit.

The different sectors of activity

Our products are also used for microbiological analysis in the food, environmental, pharmaceutical, cosmetic, animal health and public institutes sectors.